CzMA JEP Home page CZECH MEDICAL ASSOCIATION J. Ev. PURKYNĚ
Journals - Article
CzMA JEP Home page News About Assocation Publishing Division Medical Journals Searching Supplements Catalogue
 
  Česky / Czech version Klin. Biochem. Metab., 14 (35), 2006, No. 3, p. 177–180.
 
The succession of results measuring catalytic concentrations of diagnostically important enzymes: A comparison of specialized interlaboratory study with the results of a routine cycle of EQA 
Friedecký B. 

SEKK, spol. s r. o., Pardubice
 


Summary:

       Two studies on the trueness of enzyme measurement were shortly compared and explained. First, highly organized EC4 study obtained practically the same results as routinely performed EQA program despite total different approach to problem. EC4 study worked with six fully homogenous measurement systems from leading world manufacturers, while participants of EQA used very heterogenic systems which consisted of equipment, reagents and calibrators from different producers. The best trueness showed in both studies measurement of ALT and AST, significant bias was determined in gamma-GT and CK measurement. Bias and traceability of α-amylase and LD seems to be better in Czech labs than in participants of EC4 study. Use of target values as criterion in EQA and strictly requirement to use only measurement methods based on the same principles as reference IFCC procedures can be way to trueness and traceability of enzyme measurement results.

        Key words:
       

Order this issue

  BACK TO CONTENTS  
 
 
| HOME PAGE | CODE PAGE | CZECH VERSION |
©  1998 - 2008 CZECH MEDICAL ASSOCIATION J. E. PURKYNĚ
Created by: NT Servis, s.r.o., hosted by P.E.S. consulting, s.r.o.
WEBMASTER