  Česky / Czech version Klin. Biochem. Metab. 10 (31), 2002, No. 4, p. 235–237
 
Modification of HPLC Assay for 5-hydroxyindoleaceticAcid Estimationin Urine Resolved the Interference inTreatment with 5-aminosalicylic Acid 
Dastych M.1, Dastych M. jr.2, Sojková V.1, Malášková L.1 

1Oddělení klinické biochemie a hematologie FN Brno 2Interní gastroenterologická klinika FN Brno
 


Summary:

       The interference of assessment of 5-hydroxyindole-3-acetic acid in urine by the HPLC assay with electrochemical detection can be caused by the aminosalicylate treatment. The separation of the interfering peak was achieved by a change of the pH of the mobile buffer phase from the value 2.6 to 3.2.

        Key words: 5-hydroxyindoleacetic acid, HPLC, aminosalicylates, interference.
       