Determination of total homocysteine level in plasma by
liquid chromatography and tandem mass spectrometry
Humplíková S.1, Minář J.1,3, Kučerová M.1, Radina M.1, Valík D.2
1Onkologické Centrum J. G. Mendela, Nový Jičín 2Oddělení laboratorní medicíny, Masarykův onkologický ústav, Brno 3Univerzita Tomáše Bati ve Zlíně, Fakulta technologická, Chemie a technologie potravin, Zlín |
|
Summary:
Objective: To develop a new LC/MS/MS - based method for routine measurement of total homocysteine in plasma.
Material and Method: Patient samples from laboratories of Cancer Centre of J. G. Mendel were analyzed by liquid
chromatography with tandem mass spectrometry.
Results: Ion transitions m/z 135.9 → 89.8 for homocysteine and m/z 140 → 90 for internal standard – homocysteine-d4
were monitored in the positive multiple reaction-monitoring mode. Our newly developed LC/MS/MS method has the
following characteristics: retention time of homocysteine and homocystine-d8 is 1.4 min, limit of quantitation 0.5 µmol/l and
limit of detection 0.1 µmol/l. Repeatability was determined by analyzing 2 plasma samples, the variation coefficients
obtained were 0.12 % at concentration of 8.3 µmol/l and 0.71 % at concentration of 18.2 µmol/l.
Conclusion: The selective, specific and rapid method suitable for routine determination of total homocysteine in plasma
was developed.
Key words:
homocysteine, liquid chromatography, tandem mass spectrometry.
|