Summary:
To assess normal intraocular pressure values of the child population the authors
examined a group of 215 children aged 4 to 15 years with a normal ophthalmolo-
gical finding, a total of 406 eyes.
All tonometric examinations were made without general anaesthesia or other
pharmacological measures influencing the air-puff contact-free tonometer NT
1100 NIDEK, always the mean of three reliable results.
Although there were certain variations of the intraocular pressure, they were not
correlated to the child´s age after the age of 4 years.
The mean intraocular pressure in the whole group of 406 eyes was 16.08 ± 3.08 mm
Hg.
Knowledge of normal values of the IOP in children is important for diagnosis as
well as for monitoring of already verified glaucoma.
Key words:
normal intraocular pressure, child´s eye
|