Česky / Czech version | Čes. a slov. Oftal., 59, 2003, No. 6, p. 397 - 401 |
Is Accurate Assessment of the Intraocular Pressure after Refraction Operations
Possible? Černák A.1, Sisková E.2, Ďurčová T.2, Černák M.1 1Očná klinika SPAM, Bratislava, prednosta prof. MUDr. A. Černák, DrSc. 2Očná ambulancia Excimer med., Bratislava, riaditeľ MUDr. E. Sisková |
|
Summary: Objective: To evaluate how the change of thickness or curving of the cornea after
refraction operations (PRK or LASIK) influences the intraocular pressure (IOP)
during assessment with a noncontact tonometer.
Place of investigation: Ophthalmological Clinic SPAM and Ophthalmological ambulance
EXCIMER.
Material and methods: With regard to the performed refraction operation the
patientswere dividedintotwogroups.Thefirstgroupcomprised87patientswhere
PRK was performed to correct myopia. The mean myopia in this group was -5.27
D ± 2.32. The second group comprised 62 patients where LASIK surgery was
performed to correct myopia. The mean myopia in this group was -10.18 ± 4.36.
IOP was assessed by means of a noncontact tonometer before and 6 months after
surgery.
Results. In the first group a mean reduction of IOP by 4.8±2.1 mm Hg occurred.
The central thickness of the cornea declined on average by 55.69 ±28.9 micros and
the curving of the cornea diminished on average by 3.72 D ± 1.76 D.
In the second group of patients the IOP declined on average by 7.97 ± 2.04 mmHg.
The central thickness of the cornea diminished by 78.75 microns ± 42.26 and the
curving of the cornea by 4.65D ± 2.07In both groups a significant correlation was found between the reduction of IOP and the central thickness and curving of the
cornea.
Conclusion: Diminution of the thickness of the central cornea and diminution of
the curving lead to lower values of the IOP. There is a direct correlation between
thinning of the cornea and the decline of IOP.
|
|