Pressure-volume Curve:Automated Measurement of Respiratory Mechanics –A.R.M.Software
RŮŽIČKA J.1 ,NALOS M.2 ,KLEIN A.3 ,KASAL E.2 ,NOVÁK I.4
1 Biofyzikální ústav,LF UK,Plzeň ,vedoucí as.MUDr.Jiří Beneš,PhD. 2 Anesteziologicko-resuscitační klinika,FN Plzeň,přednosta doc.MUDr.Eduard Kasal CSc. 3 Alex Electronics,Plzeň 4 Metabolická jednotka intenzivní péče,Interní klinika FN Plzeň,přednosta MUDr.Ivan Novák |
|
Summary:
We developed software able to control mechanical ventilator (Adult Star TM Infrasonics,USA)via serial port and to simultaneously read data from
the ventilator output.This enables to deliver randomly selected test tidal volumes under constant flow conditions with computer controlled expiratory
and inspiratory valve holds and to record the volume and respective pressures for automated pressure-volume curve measurements.We tested the
software performance in 4 patients suffered by a several lung injury on three different PEEP levels.Data obtained automatically from the ventilator
were compared with independent calibrated measuring system consisting of airway and oesophageal pressure transducers (Validyne ±100 cmH2O,
USA),and heated pneumotachograph (Hans Rudolph,USA)connected to differential pressure transducer (Validyne ±2 cmH2O,USA).Data were
processed and collected into the same computer operating the ventilator and analyzed.
A.R.M.software driven ventilator is able to perform static respiratory mechanics measurements without disconnecting the patient from the ventilator
in 1 hour intervals.No additional equipment other than ventilator and external computer controlled end expiratory valve is needed.
Key words:
ARDS –ALI –automatization –mechanical ventilator
|