SPECT: The Use of the Jaszczak Phantom for Evaluating Tomographic Parameters
of Scintillation Cameras Nechvíl K. Oddělení lékařské fyziky, Nemocnice Na Homolce, Praha, vedoucí doc. ing. J. Novotný, CSc. |
|
Summary: The study emphasizes the importance of tomographic parameters for the quality of
SPECT (single photon emission computed tomography) image and presents the procedure
for their measurement with the use of a cylinder Jaszczak phantom. Specific procedures
for calculating individual tomographic parameters are described and the
obtained results are presented. The calculated values of individual parameters are
compared with the results of an extensive study published abroad. In conclusion, the
possibilities of implementation of recommended procedures into clinical practice are
evaluated and analyzed.
|
|