Dosimetric Verification of Treatment Planning System CadPlan Using the
AAPM Report No. 55 Wicha R.1,3, Šimíček J.2, Vrba M.2, Berkovský P.3 Varian Medical Systems, Palo Alto, CA950351 Masarykův onkologický ústav, Brno2 ředitel doc. MUDr. R. Vyzula, CSc. Zdravotně sociální fakulta Jihočeské Univerzity, České Budějovice3 |
|
Summary: Dosimetric verification of radiotherapy treatment planning system (TPS) represents an important
step in the commissioning process. The accuracy of a calculated dose distribution is one of the most
important requirements for a good commercial treatment planning system. Verification of the TPS
accuracy may be performed in two different ways – by comparing calculated and measured data for
given geometries and test points or by utilising a standard set of test data. Results of dosimetric
verification of the CadPlan treatment planning systemusingAAPMReport No. 55 data are presented
in this paper. In the case of homogenous phantom the average deviation was 0.8 % and we got a pass
result for tolerance 2% in 98% of the test points, average deviation in the heterogenous phantom was
1% and the pass rate for tolerance 3% was 100%. Results of the dosimetric verification show that
CadPlan meets commonly accepted criteria for a clinical use.
|
|