Centr. eur. J. publ. Hlth 7, 1999, No. 4, p. 197–202 | |
PROBLEMS ASSOCIATED WITH
STERILIZATION USING ETHYLENE OXIDE.
RESIDUES IN TREATED MATERIALS Buben I. 1 , Melicherčíková V. 2 , Novotná N. 1 , Svitáková R. State Institute for Drug Control, Prague 2 National Institute of Public Health, Prague 3 Institute of Macromolecular Chemistry, Academy of Sciences, Prague, Czech Republic |
|
Summary: The paper deals with problems associated with reduction of undesirable effects of ethylene oxide in polymers in medical devices on the
patient's health. The authors explain the need of careful elaboration and validation of the sterilization and aeration process incl. assessment
of ethylene oxide (EO) residues.
The authors investigated the effect of the type of material and conditions of sterilization and aeration on the assessed EO concentration.
For research of the behaviour of different polymers in the sterilization process model sterilizations of actual items of medical devices with
a known composition proved more suitable than assessment in medical devices from medical institutions. The main conclusions of the
investigation were a classification of polymers into those suitable and unsuitable for sterilization or resterilization, and attention was also
drawn to poor reproducibility of results in old sterilizers, in particular those lacking effective aeration in aerators.
|
|