Summary:
Study objective: Design and validation of a real-time PCR assay for quantitative detection of C.
pneumoniae and C. trachomatis in clinical specimens.
Material and Methods: A part of the 16S RNA gene was selected as the target sequence for
amplification. The reaction product was cloned in the bacterial plasmid and a recombinant
calibrator and a combined internal standard were prepared, usable in three different PCR
assays. Archived DNA isolates from various clinical specimens, DNA isolates from other infectious
agents and control samples for DNA detection, CT (QCMD) and CPN (Instand), were used
for validation.
Results: Reaction specificity and reproducibility were tested. The analytical sensitivity was set to
5–10 copies of the bacterial genome/reaction. As many as 105 DNA isolates from various types of
clinical specimens (swabs, urine, peripheral blood and BAL fluid) were tested by real-time PCR
and conventional PCR assays. The specimens that had not yielded concordant results were characterized
by a third independent amplification test. The sensitivity and specificity of real-time
PCR were 89 % and 100 %, respectively. The assay is suitable for both screening and diagnosis of
Chlamydia.
Key words:
Chlamydia pneumoniae – Chlamydia trachomatis – diagnosis – real-time PCR.
|