CzMA JEP Home page CZECH MEDICAL ASSOCIATION J. Ev. PURKYNĚ
Journals - Article
CzMA JEP Home page News About Assocation Publishing Division Medical Journals Searching Supplements Catalogue
 
  Česky / Czech version Čes. Radiol., 55, 2001, No. 2, p. 109-114
 
Verification of the Calculation Accuracy of Linear Accelerator Clinac 600C Enhanced Dynamic Wedge Irradiation Fields  
Šimíček J., Wicha R., Vrba M., Prchal F. 

Masarykův onkologický ústav, Brno, ředitel doc. MUDr. Jan Žaloudík, CSc. Zdravotně sociální fakulta Jihočeské univerzity, České Budějovice Varian Medical Systems, Palo Alto, CA 95035
 


Summary:

       Enhanced dynamic wedge is one of the dynamic delivery techniques in radiotherapy, which allows optimizing final dose distribution. Treatment planning the above-mentioned technique is done with the help of a treatment planning system (TPS). Calculation accuracy of commercially available TPS CadPlan has been performed for a variety of fields with enhanced dynamic wedge. Data calculated by TPS were compared with measurements in water phantom. The comparison based on percentage dose curves and transversal profiles indicated, that for all calculation points the agreement between calculation and measurement was better than 2% or 4 mm respectively.

        Key words: linear accelerator - treatment planning system - dynamic wedges
       

Order this issue

  BACK TO CONTENTS  
 
 
| HOME PAGE | CODE PAGE | CZECH VERSION |
©  1998 - 2008 CZECH MEDICAL ASSOCIATION J. E. PURKYNĚ
Created by: NT Servis, s.r.o., hosted by P.E.S. consulting, s.r.o.
WEBMASTER